Dokument
IMA 201
Rev
2
Page
1 (4)
Installation manual
CF3-W-EU, CF3-W-EU-Disp
CO
2
transmitter
CF3-W-EU CF3-W-EU - Disp
General
The IAQ-sensor product CF3-W-EU (sensor for wall mounting) is designed to
measure carbon dioxide (CO
2
) in rooms. Option - TR is prepared for temperature
measurements by the resistive temperature probe mounted by the user. The
temperature probe is potential free (floating). Option - Disp displays the measured
CO
2
value in ppm (parts-per-million) on the LCD.
The units are designed for connecting to Direct Digital Control (DDC) with 0-10V or
2-10V signal inputs. The two parallel signal outputs OUT1 (0-10V) and OUT2 (2-10V
or 4-20 mA) give linear signal voltages or currents corresponding to the measuring
range.
The output OUT2 also indicates the status by setting the output voltage to 1V or the
output current to 2 mA when the sensor self-diagnostics detects any error.
To open the wall mounted housing
Figure 1. Closed housing seen from
above. The housing is opened by
pressing a screw driver on the locking
hook. The locking hook is then
released.
Press here to open.
Page view 0
1 2 3 4

Summary of Contents

Page 1 - CF3-W-EU, CF3-W-EU-Disp

Dokument IMA 201 Rev 2 Page 1 (4) Installation manual CF3-W-EU, CF3-W-EU-Disp CO2 transmitter CF3-W-EU CF3-W-EU - D

Page 2 - Electrical connections

Dokument IMA 201 Rev 2 Page 2 (4) Electrical connections The power supply has to be connected to and . is considered as system ground

Page 3 - Maintenance

Dokument IMA 201 Rev 2 Page 3 (4) Figure 3. Enlarged picture of the PCB with the jumper set to current output (left position) Figure 4. En

Page 4 - Dimensions

Dokument IMA 201 Rev 2 Page 4 (4) Dimensions Figure 5. The dimensions of the sensor in mm and (inches) Mounting of the sensor onto the wall

Comments to this Manuals

No comments