Rotronic CP11 User Manual

Browse online or download User Manual for Tools Rotronic CP11. ROTRONIC CP11 User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 1
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
SHORT INSTRUCTION MANUAL
General description
The CP11 is an accurate hand-held datalogger that displays & records relative humid-
ity, temperature, CO2 and calculated parameters such as dew point and wet bulb.
These short instructions are limited to a description of the main functions and installa-
tion of the device. The detailed instruction manual can be found on the internet at
www.rotronic-humidity.com
Programming
Most of the settings, such as memory clear, CO2 alarm, units( C/ F), sampling rate,
pressure compensation and real time clock can be changed using the function keys.
Built-in CO2
sensor w/
calibration cap
PC connect port
5VDC adaptor port
Battery
compartment
Integrated probe
Tripod hole
Power Supply
The CP11 uses 4pcs AA alkaline batteries or
5VDC adaptor (optional accessory). To insert
(replace)the batteries, use the screw driver
enclosed to loose 2 screws and put in the batteries
in correct polarity with good contact.
Connection to PC or Laptop
Connect the CP11 via mini USB cable to a PC or
Laptop. Recorded data download can be easily
downloaded with the software enclosed.
Functions Key
ON/OFF/SETUP
-Press to turn on or turn off the instrument
-In normal mode, press longer to enter setup mode.
-Press with HOLD key to disable sleep mode
HOLD/UP
-Freezes the current readings
-Cancels data hold function
-Selects unit or increases value in setup
MODE/RECALL
-Press to switch displayed parameter
-Press longer to enter memory recall
START/ESC
-Start automatic logging mode
-Exits setup/recall/logging
MEM/DOWN
-Press to manually record the reading
-Selects unit or decreases value in setup
MIN/MAX/AVG/ENTER
-Activates MIN,MAX,AVG function
-Saves and finishes settings
Display
Upper LCD Humidity/TEMP/CO2....reading
Lower LCD Real time clock display
CO2 Carbon dioxide reading
HOLD Readings are hold unchanged
MIN/MAX Minimum/Maximum readings
AVG Average reading
Low battery indicator
DP Dew point temperature
TEMP Temperature
WBT Wet bulb temperature
%rh Unit of relative humidity
o o
C/ F Celsius/Fahrenheit of temperature
CAL In calibration status
REC In manual/automatic logging status
RECALL In manual records recall mode
CO2
%rh%rh
TEMP
DP
WBT
OO
FF
OO
CC
REC
Recall
AM
MAX
MIN
HOLD
AVG
CAL
PM
Press to turn the meter on. At power up,
it emits a short beep and per-forms 28 sec.
countdown (Fig.1) for meter warm up, then
enters normal mode with real time clock
displayed (Fig.2). Date and time are
displayed in turns.
POWER ON/OFF
SET
ON/ OFF
CO2
AM
Fig.2
TAKING MEASUREMENT
The meter starts measurement when power on and update readings every second. In
the condition of operating environment change, it takes 30 seconds to respond for
CO sensor.
NOTE: Do not hold the meter close to faces in case exhalation affects CO2
2
To power off, press again to turn off the
meter.
Operation
Fig.1
%rh%rh
TEMP
OO
CC
AM
Press to switch to temperatures display(Fig.4,5,6).
The lower display shows the real time clock.
TEMP,DP,WBT
Press to switch the mode to CO2.User can get the
CO2 reading in ppm unit on main display.(Fig.7). The
lower display shows the real time clock.
CO2 (Carbon Dioxide)
Fig.7
Fig.5
Fig.6
Press to switch to humidity display.(Fig.3).The
lower display shows the real time clock.
Humidity
Fig.3
CO2
AM
Fig.4
OO
CC
AM
DP
OO
CC
AM
WBT
MODE
RECALL
In normal display mode, press to freeze the
readings, "HOLD” icon is flashed on the left top of LCD
(Fig.8). All current readings are kept unchanged.
Press again to cancel the hold function.
DATA HOLD
Fig.8
BACKLIGHT
HOLD
NOTE:
The hold function is disabled in Min/Max/Avg mode.
START
ESC
Fig.9
Fig.10
Fig.11
Fig.12
%rh%rh
HOLD
CO2
AM
CO2
MAX
CO2
AVG
MAX/AVG
MIN
HOLD
The backlight will be activated for 10 seconds by
pressing any key.
This meter allows you to check the minimun, maximum,
average, current value from the moment you press the
key. Under normal mode, each press of displays
MIN, MAX, AVG and current value in sequence.
(Fig.9~11). To return to normal mode, press to
escape(Fig.12)
In this max/min/avg/current modes, it shows the
corresponding readings on main display and the
accumulated time of how long you are in this mode.
The maximum time interval for this function is 18 hours.
Press can switch the displayed parameters. (Fig.13)
Distinguishing the normal display and current review
by checking the time display format. In normal display,
date & time are displayed in turns. In current review, the
lower display is the accumulated time of how long you
are in max/min/avg mode.
MIN,MAX,AVG,CURRENT REVIEW
MAX/AVG
MIN
NOTE:
HOLD” & “MANUALLY RECORD” & “RECALL” functions are disabled in Min/
Max/Avg mode.
%rh%rh
MAX
CO2
ALARM
The meter features audible alarm to give warnings when CO2 concentration exceeds
the limit. (See “Meter Setup” section for setting alarm threshold). It emits beeps (Abt.
80dB) when CO2 level goes over the set value and stops only when the readings
fall below the set value. It beeps again when value goes over the limit.
The meter turns off automatically after 20 minutes of inactivity.
To override the function, first, the meter is in power off status.
Hold down and for 2 seconds to turn on the meter until
n” appears.(Fig.14)
SET
ON/ OFF
HOLD
AUTO POWER OFF
DATA LOGGING
The meter can automatically record Humidity/TEMP./CO2 for long term environment
monitor. The memory capacity is 6000 points for each parameter. Users can set up
sampling rate from 1 second to 4hours 59 minutes and 59 seconds. The factory
preset rate is 30 second.
START
ESC
REC
Fig.19
CO2
REC
AM
Repeat above to start another run of logging. While the logging
space is full, it shows “FULL. To terminate data logging, press
for 2 seconds, icon stops flashing.
Press key to switch to the mode you want to see. Except the
mode switching function, auto power off, min/max, manual
record, hold and recall functions are all disabled during logging.
REC
START
ESC
MODE
RECALL
Meter Setup
When meter is powered on , hold down for more than 2 seconds until entering
meter setup mode. To exit setup, press to return to normal mode.
Press or key to select the program and press key to select.
The programmable setting flashes on display, press or key to select and
press key to confirm. To leave without saving, press key to return.
SET
ON/ OFF
START
ESC
HOLD
Fig.13
MEM
MAX/AV G
MIN
HOLD
MEM
MAX/AV G
MIN
START
ESC
MODE
RECALL
MODE
RECALL
MODE
RECALL
Fig.14
NOTE:
Auto sleep function will be disabled during calibration &
auto recording mode. Auto sleep function WILL NOT be
disabled automatically via using adaptor.
P10, 99 points MEMORY clear
P11, choose yes or no
Note:
“ “is displayed while
memory is cleared.
P20, auto LOG data clear
P21, choose yes or no
Note:
“ “is displayed while
memory is cleared.
P30, CO2 ALARM setup
P31, choose from 100
to 9900
Note:
-Factory preset 1400ppm.
-Adjustable scale is every
100ppm
o o
P51, choose C or F
Note:
o
-Factory preset C
P50, Temp. UNIT setup
o
C
o
F
P61, choose from
04:59:59 to
00:00:01
Note:
-Factory preset at 30 sec.
-The format is
hour:minute:second
P60, logging sampling RATE
P71, choose from
700 to 1990hpa
Note:
-Factory preset at 1013hpa
-The adjustable scale is
every 1hpa.
P70, PRESSURE compensation
P81, choose 12 or 24
hour format
Note:
-Factory preset at 12H &
dd-mm-year format.
P80, REAL TIME CLOCK setup
choose DATE
format
Input Year/
Month/ Day
year
month
date
Input Hour/
Min./Sec.
Hour
Minute
Second
or
Humidity Calibration
This meter can be calibrated either via 35% & 80% salt bottle or humidity generator.
o
The ambient condition is recommended as stable 25 C. While the meter is powered
off, plug the sensor probe into 35% salt. Press , , simultaneously for 3
seconds to turn on the meter and enter calibration mode. Press or to select
35.0% calibration and press to start. “CAL” and calibrating value are blinking on
the LCD.
CAUTION:
Do not calibrate the humidity without the default calibration salt. Otherwise,
it will cause permanent damage. Contact ROTRONIC for calibration salt or
services.
Single point calibration will cause error code E11 or non accurate reading.
Always do dual point calibration to complete a process. Suggest to calibrate
low humidity first.
CO2 Calibration
CAUTION:
Do not calibrate the meter in the air with unknown CO concentration.
2
Otherwise, it will be calibrated as 0 or 400ppm by default and leads to
inaccurate measurements.
Ensure the batteries are with full voltage during the calibration to prevent from
interruption or failed calibration. When the accuracy becomes a concern after
a long time usage, return to dealers for standard calibration.
Technical Data
Humidity/accuracy: 0.1% to 99.9%/ ±3.0%(10~95%@25 °C), ±5% (other)
Trouble Shooting
Error Messages Solution
E01 CO sensor is out of order Turn off meter and re-start again
2
E33 CO sensor is out of order Retry CO calibration
2 2
E02 measured value is under range Put meter in normal condition
E03 measured value is over range Put meter in normal condition
E04 Original data error results in
this error (DP, WB)
E11 RH calibration error Retry humidity calibration
E31 Temp. sensor or AD damaged Return for repair
E32 Memory IC damaged Return for repair
E33 RH sensor or circuit damaged Return for repair
Solve the original data error
ROTRONIC AG, CH-8303 Bassersdorf
Tel. +41 44 838 11 44, www.rotronic.com
ROTRONIC Messgeräte GmbH, D-76275 Ettlingen
Tel. +49 7243 383 250, www.rotronic.de
ROTRONIC SARL, 56, F - 77183 Croissy Beaubourg
Tél. +33 1 60 95 07 10, www.rotronic.fr
ROTRONIC Italia srl, I- 20157 Milano
Tel. +39 2 39 00 71 90, www.rotronic.it
ROTRONIC Instruments (UK) Ltd, West Sussex RH10 9EE
Phone +44 1293 571000, www.rotronic.co.uk
ROTRONIC Instrument Corp, NY 11788, USA
Phone +1 631 427-3898, www.rotronic-usa.com
Singapore
Phone +65 6 , www.rotronic.com.sg
ROTRONIC Shanghai Rep. Offi ce, Shanghai 200233, China
Phone +86 40 08162018, www.rotronic.cn
ROTRONIC Instruments Pte Ltd, 159836
376 2107
o o o
Temperature/accuracy: -20...60 C/+0.3 C@5~40 C
CO2 /accuracy: 0...9999ppm/+(30ppm+5% of reading)@0~5000ppm
o
Storage and transit: -20...60 C/10...90%rh, non condensing.
o o
Operating limit at electronics:0...50 C for CO2, -20...60 C for rest/ non condensing.
Memory: 99 records manually, 18000 records auto logging.
Dimensions: 234(L)x77(W)x42(H),mm
Weight: 200gr.
Battery type: 4pcs AA alkaline
DC power type: 5VDC, 250mA. Optional available.
PC connection: mini USB port
Congratulations on your purchase of the new portable humidity, temperature and CO2
instrument. Please read these short instruction carefully before installing the device.
SET
ON/ OFF
DATE/TIME SETUP
When meter is powered on , hold down for more than 2 seconds until entering
meter setup mode. Press or key to select the program P80 (rtc) and press
key to enter.
First, choose the time format as 12hour or 24hour by pressing or and then
press to confirm. Second, choose the date format as Month-Day-Year or Day-
Month-Year by pressing or and then, press key to confirm. Now, start to
input the real time clock value from year, month, day, hour, minute to second. Press
or to select and then press to confirm. Press to return to normal.
SET
ON/ OFF
HOLD
MEM
MAX /AVG
MIN
START
ESC
HOLD
MEM
HOLD
MEM
MAX /AVG
MIN
HOLD
MEM
MAX /AVG
MIN
MAX /AVG
MIN
UNIT SETUP
When meter is powered on , hold down for more than 2 seconds until entering
meter setup mode. Press or key to select the program P50 (unit) and press
key to enter.
Press or to select or and then press to confirm. Press to return
to normal.
SET
ON/ OFF
MAX /AVG
MIN
START
ESC
HOLD
MEM
MAX /AVG
MIN
HOLD
MEM
o
C
o
F
MANUALLY RECORDING
REC
MEM
Fig.15
HOLD
99 MEMORIES RECALL
In normal or hold mode, press key for more than 2 seconds until the icon
flashes. Press or to scroll the memories. The memory sequential number
displays on main LCD first and reading comes after. (Fig16&17). Press key to
switch the display parameter.(Fig.18).The time displayed in memory recall mode is
the recording time of this memory.
Reca ll
HOLD
MEM
Fig.16
Fig.17
REC
AM
MODE
RECALL
Rec al l
CO2
Rec al l
Fig.18
Rec al l
%rh%rh
MODE
RECALL
After sampling rate is selected, press for 2 seconds under normal mode to start
logging. The icon flashes to indicate the logging status and LCD main display
shows the real time value. Lower displays are real time clock.(Fig.19)
To setup the sampling rate, when meter is powered on , hold down for more than
2 seconds until entering meter setup mode. Press or key to select the program
P60 (rAtE) and press key to enter.
Press or to select the sample rate from hour, minute to second and then press
to confirm. Press to return to normal.
SET
ON/ OFF
MAX /AVG
MIN
START
ESC
HOLD
MEM
MAX /AVG
MIN
HOLD
MEM
5VDC universal adaptor: AC1214
CO2 calibrator: CO2 CALIBRATOR
The meter features a 99 point memory. In the normal or hold mode, press key
to start recording, the icon and the main display will be flashed 3 times. The
memory sequential number will show on the main display with 99 points in maximum.
The data in the memory contains all parameters information and not only the displa-
yed parameter. When all 99 points memory are occupied, a “FULL” message will be
displayed on the screen. If you think the reading is changing too fast, you can press
key to freeze current reading first then manually record the data. This recording
function will be disabled in the Min/Max mode.
Optional Accessor y
Part Number
Batter y low indicator
A battery low icon “ “ appears when the supplied power is too low to maintain a
normal operation. Please replace the batteries for accurate measuring. Suggest
to use adaptor (optional part: AC1214) for long term measuring & datalogging.
SET
ON/ OFF
START
ESC
MEM
HOLD
MEM
MAX/AV G
MIN
P40, CO2 ABC setup
P41, Choose on or off
Note:
-Factory preset ABC
function as ON
SET
ON/ OFF
START
ESC
MEM
HOLD
MEM
MAX/AV G
MIN
The sensor has an automatic calibration function to secure long-term accuracy. If a
calibration is desired, press simultaneously for 3 seconds to turn on the
meter and enter calibration mode. Press or key to select 400 ppm CO2
calibration (Outside Air calibration). Press to start and seeCAL” and CO2 value
blinking on LCD. Wait about 10 minutes until the blinking stops and the calibration is
completed and back to normal mode. While choose 0 ppm CO2 calibration, remove
the cap from meter rear side and plug the CO2 CALIBRATOR pipe into the sensor
hole beneath. Wait about 10 min until the blinking stops and the calibration is
completed and back to normal mode.
Page view 0
1

Summary of Contents

Page 1 - SHORT INSTRUCTION MANUAL

SHORT INSTRUCTION MANUALGeneral descriptionThe CP11 is an accurate hand-held datalogger that displays & records relative humid-ity, temperature, C

Comments to this Manuals

No comments