Rotronic HF5 User Manual Page 23

  • Download
  • Add to my manuals
  • Print
  • Page
    / 30
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 22
E-M-HF5-V1_25
Rotronic AG
Bassersdorf, Switzerland
Document code
Unit
HygroFlex HF5 Humidity Temperature
Transmitters: User Guide
Instruction Manual
Document Type
Page
23 of 30
Document title
© 2009-2013; Rotronic AG E-M-HF5-V1_25
HF53 and HF55 transmitters:
Main Menu
Menu Items
Selections / Information
Notes
Device Settings
Units
Metric / English
Back Light
Key Press / On / Off
Display backlight mode
Contrast
LC display contrast adjustment
Trend
On / Off
Trend indication on the display
Device Information
Version
Firmware version
Serial Nbr
Serial number
Address
RS-485 address
Type
Device type
Name
Device name
User defined
Probe Information
Version
Firmware version
Serial Nbr
Serial number
Address
RS-485 address
Name
Device name
User defined
SensorTest
Humidity sensor status
Off / Good / SQ-Tuned / Bad
Record
On / Off
Data recording by the probe
(max. 2000 values)
Humidity Adjust
RefValue
Humidity reference value
± 0.1 %RH steps
Acquired
Number of cal. points in probe
memory
<Acquire>
Save cal. point to probe memory
<Delete>
Erases all calibration points
<Adjust>
Effect depends on number of
calibration points
Temperature Adjust
RefValue
Temperature reference value
± 0.1 ˚C steps
<Adjust>
1-point adjustment only (offset)
o Record: both the recording mode (start / stop and the log interval cannot be changed from the menu and
are as configured with the ROTRONIC HW4 software
o SensorTest: Off means that the humidity sensor has not been tested due to the configuration settings of
the test. For a description of the automatic humidity sensor test and drift compensation (SQ-tuning) see
documents E-T-AC3000-DF-V1 and E-M-HW4v3-F2-005