10 Technical data
10.1 Specifications
Humidity temperature transmitter with analog output
signals
Humidity temperature
transmitter with digital
output
Mechanical configuration types
Power supply and connections
10…28VDC
V min = 10 V + (0.02 x
Load*) *Load
resistance in ohms.
Option 1 (standard):
15…40 VDC or
12…28 VAC
Option 2: (galvanically
isolated):
9 to 36 VDC or
7 to 24 VAC
Option 1 (standard):
5…40 VDC or
12…28 VAC
Option 2: (galvanically
isolated):
9 to 36 VDC or
7 to 24 VAC
Option 3: PoE (not
available on models
with RS-485
interface)
Typical current consumption
Model with USB
interface: 100 mA
Model with Ethernet
interface: 300 mA
Terminal block and M16 cable grip or ½” conduit
adapter
Connector (USB or
Ethernet), terminal block
(power supply and RS-
485) and M16 cable grip
or ½” conduit adapter
Humidity and temperature
measurement
See document E-M-HC2 Probes > Specifications
Psychrometric calculations
Dew point (Dp) above and below freezing
Frost point (Fp) below freezing and dew point above freezing
Wet bulb temperature (Tw)
Enthalpy (H)
Vapor concentration (Dv)
Specific humidity (Q)
Mixing ratio by weight (R)
Vapor concentration at saturation (Dvs)
Vapor partial pressure (E)
Vapor saturation pressure (Ew)